Current Generator |
|
|
Current outputs |
3 x 0...32 A AC (430 VA) / 1 x 0...96 A AC (1000 VA) / 1 x 0...85 A DC (450 W) |
|
Accuracy |
typical ±0.02% of the value ±0.01% of the range; |
|
Distortion |
0.05% typical; 0.15% guaranteed |
|
Resolution |
0.1 mA at 32 A |
|
Connections |
4 mm banana sockets |
|
Voltage Generators |
|
|
Voltage outputs |
4 x 0...300 V (85 VA) / 3 x 0...300 V (100 VA) / 1 x 0…600 V (200 VA) |
|
Accuracy |
typical ±0.02% of the value ± 0.01% of the range; |
|
Distortion |
0.015% typical; 0.03% guaranteed |
|
Resolution |
0.4 mV at 12.5 V; 10 mV at 300 V |
|
Connections |
4 mm banana sockets |
|
Other Generator Characteristics |
|
|
Output Frequency |
|
|
Currents and voltages output frequency |
0 to 3000 Hz |
|
For the voltage |
3 kHz at 60 V; 2 kHz at 100 V; 700 Hz at 300 V |
|
Transient |
0 to 5000 Hz |
|
Possibility to program 12 different frequencies on all outputs |
✓ |
|
Maximum frequency error |
0.5 ppM |
|
Resolution |
< 5 μHz |
|
Phase Angle |
|
|
Range |
- 360° ... +360° |
|
Resolution |
0.001° |
|
Accuracy (voltages and currents) 50/60 Hz |
0.01° typical, 0.02° guaranteed |
|
Battery Simulator |
|
|
Range |
0…260 V DC / 1 A |
|
Power |
50 W or 1 A |
|
Accuracy |
2% |
|
Connections |
4 mm banana sockets |
|
Binary Inputs |
|
|
Number of inputs |
12 inputs |
|
Galvanic isolations |
six groups of two inputs each, with six common points isolated among them |
|
Inputs characteristics |
potential-free or with voltage, from 4.5 to 300 V DC (24 to 230 V AC) |
|
Selection of the type of input |
dry; 5 V; 24 V; 48 V; 100 V |
|
Trigger conditions |
N.O./N.C./Edge/boolean (each input independent) |
|
Timer range |
infinite |
|
Timer resolution |
0.01 ms |
|
Timer accuracy |
0.001% of the measure ± 0.1 ms |
|
Sample rate |
up to 10 kHz; with the Transcope option up to 50 kHz |
|
Connections |
4 mm banana sockets |
|
Binary Outputs Relays |
|
|
Number of binary outputs |
4, make and break |
|
Type |
potential free timed relays |
|
Characteristics of the contacts with a resistive load |
AC: 300 V; 8 A; 2400 VA |
|
Analog DC Measuring Inputs |
|
|
DC current measuring ranges |
± 20 mA and ± 5 mA |
|
Accuracy |
20 mA: ± 0.02% of value ± 0.01% of range |
|
Connections |
4 mm banana sockets |
|
DC voltage measuring range |
± 10 V |
|
Accuracy |
± 0.02% of value ± 0.01% of range |
|
Connections |
4 mm banana sockets |
|
Other Characteristics |
|
|
Type of interfaces |
USB, Ethernet, IEC 61850, IRIG-B |
|
Internal memory |
256 Mb internal memory suitable to store in the test set approximately 2.000 test results |
|
Display |
256 colours, type LED, graphic 320 x 240 pixels; dimension 5.7 inches |
|
Mains power supply |
85 to 264 V AC, sinusoidal, single phase |
|
Frequency |
45 to 65 Hz |
|
Power consumption |
stand-by: less than 150 W; |
|
Connection |
standard 16 A AC socket |
|
Dimensions without the handle |
150 (h) x 466 (w) x 423 (d) mm |
|
Weight |
18 kg |