Range |
Resolution |
Test current |
Accuracy |
|
0...999.9 μΩ |
0.1 μΩ |
10 A |
±(0.25% m.v. + 2 digits) |
|
1.0000...1.9999 mΩ |
0.0001 mΩ |
||
|
2.000...19.999 mΩ |
0.001 mΩ |
||
|
20.00...199.99 mΩ |
0.01 mΩ |
1 A |
|
|
200...999.9 mΩ |
0.1 mΩ |
0.1 A |
|
|
1.0000...1.9999 Ω |
0.0001 Ω |
||
|
2.000...19.999 Ω |
0.001 Ω |
10 mA |
|
|
20.00...199.99 Ω |
0.01 Ω |
1 mA |
|
|
200.0...1999.9 Ω |
0.1 Ω |
0.1 mA |
|
|
Technical specification |
|||
|
Insulation type |
double |
||
|
Measurement category |
III 300 V |
||
|
Ingress protection |
IP54 |
||
|
Protection against external voltage |
up to 440 V AC for 10 s |
||
|
Power supply to battery charger |
100 V…250 V/50 Hz…60 Hz, 200 mA |
||
|
Battery charging time |
ca. 2.5 hours |
||
|
Number of measurements with 10 A current |
300 |
||
|
Maximum wire resistance for 10 A current |
0.1 Ω |
||
|
Accuracy of measuring current setting |
± 10% |
||
|
Dimensions |
295 x 222 x 95 mm |
||
|
Meter weight |
1.7 kg |
||
|
Operating temperature |
0°C…+40°C |
||
|
Charger operating temperature |
+10°C…+35°C |
||
|
Storage temperature |
-20°C…+60°C |
||
|
Humidity |
20%…90% |
||
|
Reference temperature |
+23°C ± 2°C |
||
|
Reference humidity |
40%…60% |
||
|
Temperature coefficient |
±0,01% d.v./°C |
||
|
Time to AUTO-OFF |
120 s |
||
|
TFT graphic display |
192 x 64 pixels |
||
|
Interface standard |
RS-232C |
||