Output voltage |
PHG 70 |
PHG 80 |
|
VLF truesinus® |
0 – 38 kVrms |
0 – 57 kVrms |
|
VLF square wave voltage |
0 – 57 kV |
0 – 80 kV |
|
Frequency range |
0.01 – 1 Hz |
0.01 – 1 Hz |
|
DC voltage |
0 to ±70 kV |
0 to ±80 kV |
|
Max. capacitive load |
Up to 20 μF |
Up to 20 μF |
|
Resolution |
0.1 kV |
0.1 kV |
|
Accuracy |
1% |
1% |
|
Output current |
PHG 70 |
PHG 80 |
|
Output current |
10 mA @ DC 70 kV |
1.8 mA @ DC 80 kV |
|
Max. burn current |
120 mA |
120 mA |
|
Resolution |
10 μA |
10 μA |
|
Accuracy |
1% |
1% |
|
Dissipation factor measurement |
PHG 70 TD |
PHG 80 TD |
|
VLF truesinus® |
0 – 38 kVrms |
0 – 57 kVrms |
|
Load range |
≥10 nF |
|
|
Measurement range |
0.1 x 10-3 – 1,000 x 10-3 |
|
|
Accuracy |
1 x 10-4 |
|
|
Resolution |
1 x 10-6 (mean value of the dissipation factor) |
|
|
Detection and compensation of leakage currents |
Automatically via the VSE box |
|
|
Partial discharge testing |
PHG 70 TD PD |
PHG 80 TD PD |
|
VLF truesinus® |
0 – 38 kVrms |
0 – 57 kVrms |
|
Theoretical measurement range |
10 – 12,800 m (at v/2 = 80 m/μs) |
|
|
Velocity of propagation (v/2), adjustable |
50 – 120 m/μs |
|
|
Sampling rate |
100 MSamples/s (10 ns) |
|
|
PD measurement range |
1 pC – 100 nC |
|
|
Accuracy |
Approx. 1% of cable length |
|
|
Resolution |
0.1 pC / 0.1 m |
|
|
Calibrator electrical charge (pulses) |
CAL1B: 0.1 / 0.2 / 0.5 / 1 / 2 / 5 / 10 nC CAL1E: 0.5 / 1 / 2 / 5 / 10 / 20 / 50 nC |
|
|
Calibrator power supply |
9 V block battery, DIN/IEC 6F22 |
|
|
General |
||
|
Display |
TFT monitor, 15.1” |
|
|
User interface languages |
Arabic, Chinese (CN), Chinese (TW), Czech, Danish, |
|
|
Power supply |
200 – 260 V, 50/60 Hz |
|
|
Option |
100 – 140 V, 50/60 Hz (with auto transformer) |
|
|
Max. power consumption |
3,500 VA |
|
|
Ambient temperature (HV generator) |
-20 to +55°C |
|
|
Storage temperature (HV generator) |
-30 to +70°C |
|
|
Relative humidity |
> 90%, non-condensing |
|
|
Dimensions (W x H x D) (HV generator) |
Approx. 483 x 623 x 775 mm |
|
|
Weight HV generator |
Approx. 160 kg |
|
|
Weight total |
From 250 kg (depending upon equipment) |
|
|
Safety and EMC |